Metrohm 915 KF Ti-Touch User Manual
Page 357

■■■■■■■■■■■■■■■■■■■■■■
31 Appendix
915 KF Ti-Touch
■■■■■■■■
343
Assignment
Pin No.
Function
Output 13
20
0 volts / GND
14
+5 volts
15
0 volts / GND
25
Table 9
Explanation of the individual functions
Function
Explanation
Start
The current method is started at the time of acti-
vation.
t
pulse
> 100 ms
Stop
The current method is canceled (Stop) at the time
of activation.
t
pulse
> 100 ms
Quit
The current command in the determination run
will be canceled at the time of activation.
t
pulse
> 100 ms
Ready
The instrument is ready to receive a start signal.
Conditioning OK
The line is set when Conditioning with KFT titra-
tion is at OK. The line remains set until the deter-
mination is started with [START].
Determination
The instrument performs a data-generating
determination.
EOD
End of Determination.
Pulse (t
pulse
= 200 ms) after a determination.
Error
The line is set for error message display.
Warning
The line is set for warning message display.