8.4 Dosing unit
■■■■■■■■■■■■■■■■■■■■■■
54
■■■■■■■■
915 KF Ti-Touch
NOTE
These parameters are important for the correct execution of the Pre-
paring (PREP command) and Emptying (EMPTY command) functions,
because the volumes of the tubing connections are taken into account.
Dosing port 1
Port
Port to be used as dosing port 1 for the PREP and EMPTY (see Figure
10, page 56) functions.
Selection
Port 1 | Port 2 | Port 3 | Port 4
Default value
Port 1
Length
Length of the tubing.
Input range
0.0 - 999.9 cm
Default value
40.0 cm
The setting 0.0 means that this tubing will neither
be rinsed nor emptied.
Diameter
Diameter of the tubing.
Input range
0.0 - 9.9 mm
Default value
2.0 mm
Dosing port 2
Port
Port to be used as dosing port 2 for the PREP and EMPTY functions (see
Figure 10, page 56).
Selection
Port 1 | Port 2 | Port 3 | Port 4
Default value
Port 3
Length
Length of the tubing.
Input range
0.0 - 999.9 cm
Default value
0.0 cm
The setting 0.0 means that this tubing will neither
be rinsed nor emptied.