11 Device manager

915 KF Ti-Touch

■■■■■■■■

107

The following table indicates the balance type that needs to be selected
for the balance model:

Balance

Balance type

AND

AND

Mettler AB, AE, AG, AM,
AJ, PE, PM, PJ, PR, XP, XS

Mettler

Mettler AT

Mettler AT

Mettler AX, MX, UMX, PG,
AB-S, PB-S

Mettler AX

Ohaus Voyager, Explorer,
Analytical Plus

Ohaus

Precisa

Precisa

Sartorius

Sartorius

Shimadzu BX, BW

Shimadzu

11.9

USB/RS-232 adapter

List of devices: USB/RS-232 adapter

Edit COM interface Edit

As a rule, balances are equipped with a serial RS-232 interface. To con-
nect a balance, you will require the 6.2148.050 cable. You can adjust the
interface parameters in the dialog Edit device / Port parameters. These
settings must match the settings on the connected device.

NOTE

If you make changes to parameter settings in this dialog, then you must
switch the Ti-Touch off and back on in order for the changes to take
effect.