30.7 Karl Fischer titration

■■■■■■■■■■■■■■■■■■■■■■

336

■■■■■■■■

915 KF Ti-Touch

Problem

Cause

Remedy

A side reaction is taking
place.

■

Use special reagents.

■

Adjust the method (increase/decrease the
temperature, external extraction).

■

See technical literature.

The pH value is not in the
optimum range anymore.

Add buffer, see technical literature.

The titration will not
be finished.

The titration cell is leaking.

■

Check seals and septum. If necessary,
replace.

■

Replace the molecular sieve.

The minimum increment is
too low.

Define Titration rate = user and increase the
minimum volume increment (Min. incre-
ment
) (see Chapter 28.1.2, page 261).

The stop criterion is unsuit-
able.

Adjust the control parameters (see Chapter
28.1.2, page 261)
:

■

Increase the stop drift.

■

Select a short delay time.

See also: The drift becomes
greater after each titration.

The sample is over-
titrated.

The increments at the end
of the titration are too
high.

■

Define Titration rate = user and reduce
the dosing rate (Max. rate) (see Chapter
28.1.2, page 261)
.
The following experiment provides a clue
for the optimum dosing rate: During condi-
tioning, display the drift and add sample
without starting the titration. Select a value
below the highest drift as dosing rate.

■

Stir faster.

The amount of methanol in
the working medium is too
low.

■

Replace the working medium.

■

Reduce the amount of solubility promoter,
if working with solvent mixtures, see tech-
nical literature.

The electrode may be cov-
ered.

Wipe off the electrode with ethanol or a suit-
able solvent.

The solution
becomes darker
after each titration.

Replace the working medium.

The electrode may be cov-
ered.

Wipe off the electrode with ethanol or a suit-
able solvent.