4 Installation

915 KF Ti-Touch

■■■■■■■■

19

The USB Sample Processor will be recognized automatically and
entered in the device manager of the Ti-Touch.

NOTE

The plug on the controller cable is protected against accidental
disconnection by means of a pull-out protection feature. If you
wish to pull out the plug, you will first need to pull back the outer
plug sleeve marked with arrows.

3 Configure the USB Sample Processor in the device manager (see

Chapter 11.11, page 110).

4.5

Setting up the titration vessel

The tutorial for the 915 KF Ti-Touch includes a detailed description of how
to set up the Karl Fischer titration cell and what needs to be taken into
account while doing so.

4.6

Connecting sensors

4.6.1

General

The measuring interface includes one measuring input (Pol.) for a polariz-
able electrode and one measuring input (Temp.) for a temperature sensor
(Pt1000 or NTC).

4.6.2

Connecting a polarizable electrode

Connect the polarizable electrode as follows:

1 Plug the electrode plug into the Pol. socket of the Ti-Touch.