27 Manual control
915 KF Ti-Touch
■■■■■■■■
243
Continuous dosing is paused with [Hold]. The label changes to
[Continue]. This can be used to continue the run.
3 Fill the cylinder
With the default settings (see "Fill automatic.", page 244) the dos-
ing cylinder is filled automatically.
Otherwise:
■
Tap on [Fill].
The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.
Parameter description
Volume
Volume which is dosed.
Input range
0.00000 - 99999.9 mL
Default value
0.10000 mL
Dosing rate
Rate at which it is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 31, page 338).
Filling rate
Rate at which the dosing cylinder is filled.