Metrohm 915 KF Ti-Touch User Manual
Page 151

■■■■■■■■■■■■■■■■■■■■■■
14 Common variables
915 KF Ti-Touch
■■■■■■■■
137
Document message
In the determination data it will be documented, that the time interval
has been expired.
Cancel determination
The determination is stopped.
14.4
Assigning a result automatically to a common varia-
ble
NOTE
This instruction is based on the assumption that the method contains a
calculation command with a calculation.
Proceed as follows to assign a result to a common variable:
1 Open the editing dialog of the result
■
In the command list, select the command CALC.
■
Tap on the button [Edit command].
■
Select the result whose value is to be assigned to a common vari-
able and tap on [Edit].
The editing dialog of the result is displayed.
2 Define result options
■
Tap on the button [Result options].
■
Activate the parameter Save as common variable.
■
Define Variable = CV01…CV25.