Chapter 4 Function Settings

4

1

Press the key corresponding to the
type of tolerance to be set.
The corresponding tolerance item
flashes.

The following display flashes.

Value

Tolerance display

HI set value

HH set value

LO set value

LL set value

HI

GO

HI

GO

GO

LO

GO

LO

HI

GO

LO

2

Press the

or

key and set the

tolerance.

3

Press the flashing tolerance key

again.
The input value is entered.

PROG No.

PRINT

AREA

DISP

ESC

G

C

ENT

HOLD

ZERO

FOCUS

A

POSITION

HI

GO

LO

Press these keys to set the value.

Hold Value Elimination

If the target has dirt or dust, the measured value may be wide of the accurate value. The
LS-7000 Series automatically detects and eliminates such values.

Hold Value Elimination Conditions
Hold values are eliminated under the following conditions.

In Tolerance Mode with Reference Value Used
• The measured value is below the reference value + lower limit hold value (i.e., Measured

value < (Reference value + Lower limit hold value)

• The measured value is above the reference value + upper limit hold value (i.e., Refer-

ence value + Upper limit hold value) < Measured value

In Tolerance Mode with Threshold Used
• The measured value is below the lower limit hold value (i.e., Measured value < Lower

limit hold value)

• The measured value is above the upper limit hold value (i.e., Upper limit hold value <

Measured value)

Note:

• Hold values are eliminated before measured values are averaged.
• The previous measured value will be kept on hold and displayed when the

present hold value is eliminated.