4-36

Chapter 4 Function Settings

4

Peak Hold

• Peak hold:

The maximum value during the sampling period specified by
external timing input is displayed and output as shown below.

• Bottom hold:

The minimum value during the sampling period specified is
displayed and output.

• Peak-to-peak hold: The maximum-to-minimum value during the period specified is

displayed and output.

• Average hold:

The average value during the sampling period specified is dis-
played and output.

Peak hold measured value

Internal
measured value

Judgment

Measured value

Measured value

Measured value

Measured value

Measured value

Measured value

Judgment

Judgment

t

Sampling

period

Sampling period

Sampling period

RESET

processing

ON

OFF

Timing input

Waiting for judgment

(---------- displayed)

Measured value

ON

OFF

RESET input

ON

OFF

Comparator output

ON

OFF

ON

OFF

BCD output

RS-232C output

Judgment awaiting output

(Comparator waiting output)

Strobe output

• Timing input is OFF during sampling periods. The maximum value during the sampling period immediately

before timing input will be output when the timing input turns ON.

• In the following cases, the monitor will display the result waiting status “---.-----” until the first measured

value is determined.

The LS-7000 is turned on, setting changes are made, program number changes are made, RESET
input is ON, or auto zero is input.

• Comparator output turns ON according to the measured value.
• In the above chart, the ON status of each output means that NPN open collector output is ON, provided

that it is normally open type. If it is normally closed type, the signal will be inverted.

• The ON status of each input means that the input is short-circuited to the COM terminal.
• RS-232C output turns ON in synchronization with timing input if the D-SEND item as an environment

setting is set to S1 or S2 only. Refer to Data Transmission Mode on page 5-9 for environment settings in
detail.

• If timing input is ON during reset processing, the monitor will display the result waiting status

“---.-----.”

• If a workpiece is eliminated during a sampling period and it becomes a

period without the internal measured value obtained, the maximum value of
the rest of the normal measured values will be output.

• Timing input and RESET input can be under RS-232C control. Refer to

page 7-8 for details.

• Strobe output time can be changed. Refer to Strobe Output Time on page 4-

48 for details.

Tips

Note:

When the measurement mode is set to Average Hold, set the averaging
number to 1.