Setting method – KEYENCE LS-7000 User Manual

Page 97

Advertising
background image

4-45

Chapter 4 Function Settings

4

The following table shows the relationship of scaling values along with their output
ranges and resolutions.

Scaling value [

μm/V]

Output range

Resolution (

)

1

±10μm

0.01

μm

2

±20μm

0.01

μm

5

±50μm

0.02

μm

10

±100μm

0.04

μm

20

±200μm

0.08

μm

50

±500μm

0.2

μm

100

±1mm

0.4

μm

200

±2mm

0.8

μm

500

±5mm

2

μm

1000

±10mm

4

μm

2000

±20mm

8

μm

5000

±50mm

20

μm

10000

±100mm

40

μm

❈If the minimum display unit selected is larger than the resolution selected, the mini-

mum display unit will be used as applicable resolution.

Setting Method

1

Press the

ENT

key when the output

setting is displayed.

2

Press the

FUNC

key four times.

Measured value

0V

+10.5V

+10V

Voltage

–10V

0

30

–10.5V

(Positive value
over range)

(Negative value
below range)

+10V :30mm
–10V : 0mm

HI

GO

LO

3

Press the

or

key and set the scaling value.

4

Press the

ENT

key.

The input item is entered and the output setting display appears again.

Tolerance Mode with Threshold
Set the measurement values to be output at
10 V and –10 V respectively so that the
scaling value will be decided. The voltage
can be changed while the result waiting
status is “---.-----.” (See Page 4-47)