Two-point calibration settings – KEYENCE LS-7000 User Manual

Page 77

Advertising
background image

4-25

Chapter 4 Function Settings

4

Two-point Calibration Settings

Make two-point calibration settings as explained below.

Setting Range
Two-point calibration requires two types of master workpieces that are different to each
other in size. If the master workpieces are target 1 (T1) and target 2 (T2) respectively,
make settings so that the following conditions will be satisfied.
T1 < T2
Furthermore, set the calibration span between 0.5 and 2.0.

Note:

In the case of two-point calibration, the number of averaging measurements is
fixed at 4,096 times.

1

Press the

ENT

key for calibration

setting.
The AREA1 calibration display appears.

HI

GO

LO

2

Set the master workpiece, monitor the measured value, and input the T1 value.

PROG No.

PRINT

AREA

DISP

ESC

G

C

ENT

HOLD

ZERO

FOCUS

A

POSITION

HI

GO

LO

Press these keys to set
the value.

Displays the
measured value.

The input part
flashes.

Indicates that
the T1 value is
displayed.

Indicates that AREA1 is selected.

Press this key to set the value
to a positive or negative one.

3

Press the

ENT

key.

4

Set the other master workpiece, monitor the measured value, and input the T2
value.

5

Press the

ENT

key.

The input item is entered and the calibration setting display appears again.

Note:

If the calibration conditions are not satisfied, “Error-1” will appear on the main
display. Reset the error status by pressing the

ENT

key or

ESC

key and make

the calibration settings again.