KEYENCE LS-7000 User Manual

Page 76

Advertising
background image

4-24

Chapter 4 Function Settings

4

1

Press the

ENT

key for calibration

setting.
The AREA1 calibration display appears.
Press the

FUNC

key to make AREA2

settings.

2

Input T1-A Value.

PROG No.

PRINT

AREA

DISP

ESC

G

C

ENT

HOLD

ZERO

FOCUS

A

POSITION

HI

GO

LO

Press these keys to set
the value.

The input part
flashes.

Indicates that
the T1-A value
is displayed.

Indicates that AREA1
is selected.

Press this key to set
the value to a
positive or negative
one.

3

Press the

ENT

key.

The input value is entered and the value in the next line flashes.

4

Input T1-B Value.

5

Press the

ENT

key.

6

Input T2-A Value.

7

Press the

ENT

key.

The input value is entered and the value in the next line flashes.

8

Input T2-B Value.

9

Press the

ENT

key.

The input item is entered and the calibration setting display appears again.

HI

GO

LO

HI

GO

LO

The input
part flashes.

Note:

If the calibration conditions are not satisfied, “Error-1” will appear on the main
display. Reset the error status by pressing the

ENT

key or

ESC

key and make

the calibration settings again.