Area check, Setting method, Chapter 4 function settings – KEYENCE LS-7000 User Manual

Page 67: Select the area to be set and press the, Press the, Key twice

Advertising
background image

4-15

Chapter 4 Function Settings

4

TIM

LO

GO

HI

OUT

OUT

FOCUS

2

1

A R E A

POSITION

HI

GO

LO

HI

GO

LO

PROG No.

PRINT

AREA

DISP

ESC

PROG

FUNC

ENT

H

Z

FOCUS

2

1

A R E A

POSITION

Area Check

Set the total number of edges to be measured. If the number of edges detected does not
coincide with the set value, the measured value will be invalidated.
This function makes it possible to prevent the measurement of abnormal values that may
result if dust or oil is stuck on the measuring head.

The monitor will display the following data if the
measured value is invalidated.

Setting Range
The setting range is between 2 and 127. Set the value to OFF if no area check is made.

Tips

The result of area checking can be external output as function output. Refer to
Selection of Function Output on page 4-17 for details.

Setting Method

1

Select the area to be set and press

the

ENT

key.

2

Press the

FUNC

key twice.

3

Press the

or

key and make

ON/OFF settings.

ON Setting

(1) Input the number of edges to be measured.

Press these keys
to set the value.

The setting part
flashes.

PROG No

AREA

DISP

HOLD

ZERO

LO

4

Press the

ENT

key after the value is set.

The input value is entered and the area display appears again.