Calibration Type Selection

The following section provides information on how to set the calibration type for each area.

1

Press the

ENT

key for calibration

setting.

2

Press the

FUNC

key twice.

Press the

FUNC

key three times to set the

AREA2 calibration type.

HI

GO

LO

3

Press the

or

key and select

the AREA1 calibration type.

PROG No.

PRINT

AREA

DISP

ESC

PROG

FUNC

ENT

H

ZE

FOCUS

2

1

A R E A

POSITION

G

L

The following items are displayed for
setting.

Logic calibration

Two-point calibration

One-point calibration

LO

LO

LO

4

Press the

ENT

key.

The input item is entered and the calibration setting display appears again.

Logic Calibration Settings

Make logic calibration settings as explained below.

Setting Range
Logic calibration does not require master workpieces. Input the non-calibrated values and
calibrated values of two measuring points (target 1 and target 2). Logic calibration uses the
following values
T1-A: Target 1 display value before calibration
T1-B: Target 1 display value after calibration
T2-A: Target 2 display value before calibration
T2-B: Target 2 display value after calibration

Make settings so that the following conditions will be satisfied.
T1-A < T2-A and T1-B < T2-B
Furthermore, set the calibration span between 0.5 and 2.0.