Setting method – KEYENCE LS-7000 User Manual
Page 84

4-32
Chapter 4 Function Settings
4
Example: If the averaging number is eight, the LS-7000 Series will output the data as
shown below.
1
Internal data
(2,400 samples/s)
Output data
2 3 4 5 6 7 8 9 10 11 12 13 14
Refresh
cycle
Refresh
cycle
A
B
C
Measurement averaging time
Measurement averaging time
Data samples 1 through 8 are
averaged, and data A is output.
Data samples 3 through 10 are
averaged, and data B is output.
Data samples 5 through 12 are
averaged, and data C is output.
Measurement averaging time
■ Setting Method
1
Press the
ENT
key when the output
setting is displayed.
2
Press the
FUNC
key.
3
Press the
or
key and set the
averaging number.
HI
GO
LO
PROG No.
AREA
DISP
ESC
PROG
FUNC
ENT
H
ZE
FOCUS
2
1
A R E A
POSITION
G
L
Averaging times
4
Press the
ENT
key.
The input item is entered and the output setting display appears again.
Note:
When the measurement mode is set to Average Hold, set the averaging
number to 1.