7-24

Chapter 7 RS-232C

7

Timing Chart
The following timing chart applies to timing input and RS-232C data output (the meas-
uring mode is not set to sample hold 2 or self-timing hold.)

TIMING input

Measured value

3.5ms

RS-232C data output

ON

OFF

• If the measuring mode is set to normal, auto peak hold, auto bottom hold, or auto

peak-to-peak hold, the measured value will be transmitted at the moment timing input
is turned ON.

• The measured value will be output when timing input turns ON if the measuring mode

is set to peak hold, bottom hold, peak-to-peak average hold, or sample hold 1.

If the measuring mode is set to sample hold 2, the following timing chart will apply.

TIMING input

Measured value

Sampling period + 3.5 ms

RS-232C data output

ON
OFF

• The measured value will be output when sampling period plus 3.5 ms elapses after

timing input turns ON. Refer to page 4-39 for the sampling period in detail.

If the measuring mode is set to self-timing hold, the measured value will be transmit-
ted when the measured value is determined.

Note:

If both OUT1 and OUT2 are set to sample hold 2 or either of them is set to
sample hold 2 while SYNC is set as a T-MODE option setting, measured
values will be determined asynchronously. At that time, if timing input turns
ON, both OUT1 and OUT2 measured values with a comma and a delimiter
will be returned when the OUT1 and OUT2 measured values are determined.

Output Format

Measured Value
The following command format will be used if ASYNC is set as a T-MODE option set-
ting.

STX

ETX

S S S S S S

S S

S

OUT1 or OUT2 measured value

Measured Value Format (See page 7-7)

Simultaneous Measured Value
The following command format will be used if SYNC is set as a T-MODE option setting.

STX S S S S S S

S S ,

S

ETX

S S S S S S

S S

S

OUT1 measured value

Measured Value Format (See page 7-7)

OUT2 Measured Value

Measured Value Format (See page 7-7)