Chapter 4 Function Settings

4

Measuring mode

The Measuring mode allows the LS-7000 Series to capture and hold measured values.

Setting Range
The LS-7000 Series has 11 types of Measuring modes. Refer to the following informa-
tion for the outline of each mode.

Normal

Sampling time

Measured value

Max. value

The measured
value is displayed
and output as it
occurs.

Sampling time

Measured
value

Minimum value

Sampling time

Measured value

Maximum-to-
minimum value

Max. value

Minimum value

Maximum-to-
minimum value

Peak Hold

The maximum
value during the
period specified
is displayed and
output.

Bottom Hold

The minimum
value during the
period specified
is displayed and
output.

Peak-to-peak Hold

The maximum-to-
minimum value
during the period
specified is
displayed and
output.

Average Hold

The average
value during the
period specified
is displayed and
output.

Auto Peak Hold

The maximum
value measured
is displayed and
output as it
occurs.

Auto Bottom Hold

The minimum
value measured
in displayed and
output as it
occurs.

Auto Peak-to-peak Hold

The maximum-to-
minimum value
measured is
displayed and
output as it occurs.

Sample Hold 1

The instantaneous
value at the time
specified is
displayed and
output.

Sample Hold 2

The value
measured during
the period specified
is displayed and
output.

Self-timing Hold

The measured value is displayed and
output with no external timing input (i.e.,
timing input). The self-timing period is
adjustable between 1 and 9,999 ms.

Self-timing value

Sampling time

ST
time

Workpiece insertion

Synchronous input
(timing input)

Sample Hold 1
value

Synchronous input
(timing input)

Sample hold 2
value

Sampling time

Sampling time

Measured value

Average
value