Comparator output setting, Output waiting voltage – KEYENCE LS-7000 User Manual

Page 99

Advertising
background image

4-47

Chapter 4 Function Settings

4

3

Press the

or

key and set the

analog update mode.

4

Press the

ENT

key.

The input item is entered and the output
setting display appears again.

Output Waiting Voltage

Take the following steps to set the voltage to be output when the display shows
“- - - - - - - -.”

3

Press the

or

key and set the

voltage.

4

Press the

ENT

key.

The input item is entered and the output
setting display appears again.

Comparator Output Setting

Adjust the comparator output settings.

Setting Range
The following comparator output settings are available.
• Strobe output time: Select the ON-time of the strobe output (i.e., one-shot output

time) from 0.4, 0.8, 1.6, 3.3, 6.6, 13.3, and 26.6 ms.

• Output logic:

Set the logic of the comparator output and WAITING output to NO
(normally open) or NC (normally closed).

• OFF-delay time:

Select the OFF-delay time of the comparator output from 0.0, 0.4,
0.8, 1.6, 3.3, 6.6, 13.3 and 26.6 ms.