8-6

Chapter 8 Specifications

8

Response Deley Time

Measurement Response Time

The following table lists the response time between the moment the dimensional
change of a target is detected and the measured result is obtained.

Averaging times

1

2

4

8

16

32

64

128

256

512

1024

2048

4096

Measurement response time (ms) Measurement averaging time (ms)

Refresh cycle (ms)

Internal processing time (ms)

3.34

4.16

5.00

6.66

10.00

16.66

30.00

56.66

110.84

219.16

435.84

869.16

1735.84

0.42

0.83

1.67

3.33

6.67

13.33

26.67

53.33

106.67

213.33

426.67

853.33

1706.67

0.42

0.83

0.83

0.83

0.83

0.83

0.83

0.83

1.67

3.33

6.67

13.33

26.67

2.5

2.5

2.5

2.5

2.5

2.5

2.5

2.5

2.5

2.5

2.5

2.5

2.5

The response time is determined by the corresponding measurement averaging time,
refresh cycle, and internal processing time (2.5 ms).

❈ If the RUN mode is set to normal, auto peak, auto bottom, or auto peak-to-peak, the

response time will be the response time of the comparator output (HH, HI, GO, LO,
LL, WAITING, or STROBE) and BCD output.

❈ The response time of analog output will be the response time of comparator output

plus approximately 0.4 ms.

❈ If the RUN mode is set to peak, bottom, peak-to-peak, average, or sample hold 1,

the response time will be the response time of the internal measured value. Refer to
pages 4-36 to 4-40.

❈ If the RUN mode is sample hold 2 or self-timing, the measured value will be sampled

with a delay of internal processing time.

Function Output Response Time

Focus

Area check

Difference check

110.84ms

3.34ms

3.34ms

Focus Characteristics

The following table shows typical examples of the ranges where the FOCUS display or
the FOCUS output turns on, provided that the focal point is at the center of the
measurement area.

LS-7030

Approx.

±12.5mm

Approx.

±10mm

Always set to OFF

LS-7070

Approx.

±40mm

Approx.

±25mm

Always set to OFF

FOCUS set value

9

5

0

LS-7010

Approx.

±0.8mm

Approx.

±0.5mm

Always set to OFF