Chapter 9 Expansion I/O Board

101

9

9.1.2 I/O Terminal Operations

1

1

1

1

1 7-level comparator output (terminals 1 to 14)
The LS-5000 series performs comparator output (

➮ See p. 35) for the output

number specified during the I/O slot setup (

➮ See p. 59.) procedure.

Note: When Pitch mode is selected, the comparator output is as follows:

HI3: Turns ON when the target’s actual number of pitches is different from the

preset number of pitches.

HI2: Unused
HI1: Turns ON when the measured value exceeds the preset upper/lower limits.
GO: Turns ON when the measured value is within the tolerance range.
LO1: Unused
LO2: Unused
LO3: Unused

2

2

2

2

2 Self-timing output (terminals 15 and 16)
In self-timing mode, the output specified during the I/O slot setup procedure turns
ON when the self-timing signal turns ON. The self-timing output is a one-shot
signal, as shown below:
➮ See p. 59 for I/O slot setup.

3

3

3

3

3 Timing input (terminals 19 and 20)
When this terminal and the GND terminal are short-circuited, the LS-5000 series
performs the measuring mode operation for the output number specified during the
I/O slot setup procedure.
➮ See p.46 for measuring mode and p.59 for I/O slot setup.

4

4

4

4

4 Reset input (terminals 21 and 22)
When this terminal and the GND terminal are short-circuited, the LS-5000 series
performs the measuring mode operation for the output number specified during the
I/O slot setup procedure.
➮ See p.46 for measuring mode and p.59 for I/O slot setup.

5

5

5

5

5 Auto-zero input (terminals 23 and 22)
When this terminal and the GND terminal are short-circuited, the LS-5000 series
instantaneously resets the measured value to “0.00000” for the output number
specified during the I/O slot setup procedure.
➮ See p.59 for I/O slot setup.

6

6

6

6

6 Analog output
The LS-5000 series performs analog output in the range of ±10 V for the output
number specified during the I/O slot setup procedure.
➮ See p.57 for analog output and p.59 for I/O slot setup.

Response delay time

Activation of self-timing
function

50 µs