Chapter 5 Outline of Measurements

24

5

5.3

Pitch Mode

➮ For the setting procedure, see p. 72.

Pitch mode enables pitch measurement using connector pins or IC leads.

This mode measures each center-to-center pitch, and displays the maximum and
minimum values.

Note 1: When Pitch mode is selected, only one measurement can be performed
with one scanning head.

Note 2: When Pitch mode is selected, the comparator output is activated as shown
below:
HI3 = Alarm output
HI2 = Disabled
HI1 = Output when the measured value exceeds the upper/lower limit.
GO = Output when the measured value is within the upper/lower limit.
LO1: Disabled
LO2: Disabled
LO3: Disabled

The analog output and RS-232C output are activated as shown below.
Analog output: Held at +10V.
RS-232C output: Output according to the command.

➮ See p. 97.

LASER ON

DISPLAY DATA

PITCH

P. 1

HEAD1

OK

Max

Min

0.5010

0.4990

HEAD2

OK

Max

Min

0.6010

0.5990

HEAD3

OK

Max

Min

0.7020

0.6980

HEAD4

OK

Max

Min

0.8020

0.7980

SCREEN

DISPLAY

DATA

LIMIT

SETUP

PROGRAM

OPTIONS

DISPLAY

DATA

Pitch

Optical axis