Chapter 6 Normal Mode
53
6
1
Edge No.
2
1
Edge No.
2
3
4
Target recognition method by measuring mode
•
DIA mode: A target is recognized when the number of edges in the measuring
area is two or more.
The target has not yet been recog-
nized.
The LS-5001 recognizes that a
target has been placed in the
measuring area when the number of
edges in the measuring area is two
or more.
•
SEG (m, n) mode: A target is recognized when a specified number of edges is
generated in the measuring area.
When SEG(2,3) is set:
The target has not yet been recog-
nized because the specified number
of edges (Edge No.) is not present in
the measuring area.
The target has been recognized
when the specified number of edges
is generated in the measuring area.