Chapter 6 Normal Mode

48

6

Note 1: The self-timing setting is effective only when self-timing mode is selected.

Note 2: The self-timing interval should be set longer than the averaging time.

Note 3: Self-timing mode is enabled only when DIA or SEG(m,n) is selected.
(When T.EDGE or B.EDGE is selected, self-timing mode is disabled.)

Normal mode
The measurement value/comparator output can be displayed/output at any time.

Timing Diagram

Peak hold mode
The maximum value during a specified period can be measured.

The sampling period is the duration between the falling edge and the next rising
edge of the synchronous input.

Timing Diagram

Note: When the reset input terminal and the GND terminal are short-circuited, the
retained value is reset and measurement starts again.

* When the synchronous input is ON, the displayed value/comparator output is retained.

Preset upper limit

Displayed value

Preset lower limit

HI OUT

ON

OFF

ON

OFF

ON

OFF

ON

OFF

GO OUT

LO OUT

Synchronous

input

(t)

Internal
measurement

Measured

value

ON

OFF

Synchronous

input

Power-ON or
RESET input

(t)

After the power-ON or RESET input is activated,
“-----” is displayed until the synchronous input is turned ON.

Peak hold value

Sampling

period

Sampling

period

Sampling

period