2 limit setup, 22 limit setup, Chapter 7 pitch mode – KEYENCE LS-5000 User Manual

Page 81

Advertising
background image

Chapter 7 Pitch Mode

75

7

2

2

2

2

2 Limit Setup

The reference value, upper limit, and lower limit can be set separately for each of
the four scanning heads.

■ Setting Procedure
1. Select the LIMIT SETUP screen.

Press the [LIMIT SETUP] button.

2. Set tolerance limits.

Press the button corresponding to the output for which tolerance limits are to be
set.

Set tolerance limits using the numeric keys, and press [ENTER].
Press [ESCAPE] to complete the entry.
Press [CANCEL] to cancel the entry. The value returns to the previous value.

Note: When the measured value exceeds the tolerance range, the value is dis-
played in reverse. The differentiation result relative to the tolerance limits is output
from the display unit and the control I/O board (optional).

When the measured value exceeds the upper limit or falls below the lower limit, the
HI1 output is turned on.

In PITCH mode, the Upper-Lower Limit indicates the difference from the Reference
value. If a measured value of more than 1.1000 mm is to be checked and the
Reference value is set to 1.000 mm, the Upper Limit should be +0.1000. If 0.9000
mm is to be checked, the Lower Limit should be -0.1000 mm.

HEAD1

• • •

HEAD4

LIMIT

SETUP

LASER ON

LIMIT SETUP

P. 1

PITCH

DISPLAY

DATA

LIMIT

SETUP

PROGRAM

OPTIONS

HEAD1

HEAD2

HEAD3

HEAD4

Select Head to set the limits.

LIMIT SETUP

HEAD1

LASER ON

P. 1

PITCH

8

7

5

6

4

2

3

1

0

9

ENTER

CANCEL

ESCAPE

+0.0100

1.0000

UPPER LIMIT:

REFERENCE:

LOWER LIMIT:

-0.0100