Chapter 6 Normal Mode

49

6

Bottom hold mode
The minimum value during a specified period can be measured.

The sampling period is the duration between the falling edge and the next rising
edge of the synchronous input.

Timing Diagram

Difference mode
The difference between the maximum and minimum values during a specified
period can be measured.

Timing Diagram

Note: When the reset input terminal and the GND terminal are short-circuited, the
retained value is reset to the present value and measurement starts again.

ON

OFF

Synchronous

input

Power-ON or
RESET input

(t)

After the power-ON or RESET input is activated,
“-----” is displayed until the synchronous input is turned ON.

Bottom hold
displayed value

Sampling

period

Sampling

period

Sampling

period

Internal
measurement

Measured

value

ON

OFF

Synchronous

input

Measured

value

Power-ON or
RESET input

(t)

After the power-ON or RESET input is activated,
“-----” is displayed until the synchronous input is
turned ON.

Difference hold
displayed value

Sampling

period

Sampling

period

Internal
measurement

b

b'

P

P'

(P-b)

(P'-b')