5 [calib] calibration setup, 55 [calib] calibration setup, Chapter 6 normal mode – KEYENCE LS-5000 User Manual

Page 42

Advertising
background image

Chapter 6 Normal Mode

36

6

5

5

5

5

5 [CALIB] Calibration Setup

Minute measurement fluctuations caused by a change in the target surface condi-
tion or target inclination can be corrected with each AREA. (

➮ See p.39 for setting

AREA.)

Setting Procedure
1. Select the CALIBRATE AREA screen.

Press the [CALIB] button. The screen shown below is displayed.

2. Press the button corresponding to the area for which you wish to perform

calibration.

Press the [RESET] button to reset the calibration. The calibration value is reset
to the factory setting.

Press the [CALIBRATE] button to go to the CALIBRATE screen.

3. Insert sample workpiece 1 to check the measured value.

Enter the value of sample workpiece 1 using the numeric keys, and accept the
value by pressing the [ENTER] button.

AREA1

CALIB

• • •

AREA4

CALIBRATE

AREA2

LASER ON

NOR

MAL

P. 1

DATA

54.21595

ESC

CALIBRATE

RESET

LASER ON

CALIBRATE AREA

NOR

MAL

P. 1

DISPLAY

DATA

LIMIT

SETUP

PROGRAM

CALIB

OPTIONS

AREA1

AREA2

AREA3

AREA4

Select Segment No. to calibrate.

CALIBRATE

AREA1

LASER ON

NOR

MAL

P. 1

8

7

5

6

4

2

3

1

.

0

9

ENTER

CANCEL

ESCAPE

TARGET1

0.0000

DATA

54.21595

TARGET1

0.0000

Enter calibration
values for targets
1 and 2, and press
OK.

OK