Chapter 6 Normal Mode

52

6

Self-timing mode
Measurement is automatically performed according to the internal synchronous
signal.

Timing Diagram

Note 1: When the self-timing hold mode is used, the external synchronous input is
ignored.

Note 2: When the reset input terminal and the GND terminal are short-circuited,
the retained value is reset and measurement starts again.

Target recognition method in self-timing mode
The LS-5000 series’ self-timing mode takes a specified duration to start sample-
holding after the LS-5001 has recognized that a target has been placed in the
measuring area. (The target recognition method varies depending on the
SEGMENT.)

The LS-5001 can measure a target regardless of the direction it enters the
measuring area from.

Internal measurement

Self-timing value

No target

No target

Target enters the

measuring area

Target goes out

of the measur-

ing area

Preset

self-timing

interval

Preset

self-timing

interval

Internal
synchronous
signal

Measured
value

ON

(t)

OFF

After the power-ON or RESET input is activated,
“-----” is displayed until the internal synchronous signal is turned ON.

Target enters the

measuring area

Target goes out

of the measur-

ing area