KEYENCE N-400 User Manual
Page 58
Chapter 3 N-400 Setup Procedure
50
3
Reading the current settings of the N-400
To read the current settings of the N-400, perform the following procedure. (This
procedure is almost the same as the procedure for sending settings.)
1. Press and hold the N-400 test switch for five seconds. When the N-400 displays
“S0 (
50
)”, the communication parameters are temporarily set as follows:
•
Baud rate: 9600 bits/s
•
Data length: 7 bits
•
Parity: Even
•
Step bit length: 1 bit
•
PLC link: Disabled
If you know the current communication parameter settings of the N-400, step 1. is
not necessary. You can send data to the N-400 by setting the communication
parameters of the host computer using [COM PORT] so that they conform to the
current settings of the N-400. However, if PLC link is enabled, you cannot send
data to the N-400 without performing step 1., because the handshaking protocol is
set for PLC link only.
2. Click on [COM PORT] to set the communication parameters of the host com-
puter according to the values listed in step 1. Select an RS-232C port and click
on [OK].
3. Click on [Send setting (Save in N-400 EEP-ROM)].
When you click on [Read N-400 settings], the settings of the N-400 can be read
on the host computer.
Tips