Source[1..n]:current:accbase – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 100

SCPI Command Reference
94
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:CURRent:ACCBase?
Description
This query returns the ACC base current. For this
query to function properly, the instrument must
be in High Wavelength Stability or On/Off
modulation mode and the SOURce[1..n]:STABle?
query must return 1 (stable).
*RST has no effect on this command.
Syntax
:SOURce[1..n]:CURRent:ACCBase?
Parameter(s)
None
Response Syntax
<ACCBase>
Response(s)
ACCBase:
The response data syntax for <ACCBase> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
The <ACCBase> response is a value
representing the ACC base current in amperes.
Example(s)
SOUR:PULM:STATE OFF
SOUR:WAV:MODE HRES
/* Under FDA jurisdiction, a password must be
entered before turning the laser on. */
SOUR:POW:PROT:SPWD "safekey"
SOUR:POW:STAT ON