:SOURce[1..n]:POWer:STATe
Description
This command turns the optical source on or off.
When source is on, the red LED (Active) on the
front of the instrument lights up.
*RST sets the optical source to OFF.
Syntax
:SOURce[1..n]:POWer:STATe<wsp>
<PowerState>
Parameter(s)
PowerState:
The program data syntax for <PowerState> is
defined as a <Boolean Program Data> element.
The <PowerState> special forms ON and OFF
are accepted on input for increased readability.
ON corresponds to 1 and OFF corresponds to 0.
The <PowerState> parameter is the new power
state of the source.
1 or ON, turns the optical source on.
0 or OFF, turns the optical source off.
Example(s)
SOUR:POW:STAT ON
Notes
When the user is under Food and Drug
Administration (FDA) jurisdiction, a password
must be entered before turning the laser on.
See Also