Source[1..n]:pulm:internal:depth – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 126

SCPI Command Reference
120
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:PULM:INTernal:DEPTh
Description
This command is used to set the dither
modulation signal depth. If the instrument is not
in Dither modulation mode, this command will
have no effect.
At *RST, the dither depth is 0.005 amperes.
Syntax
:SOURce[1..n]:PULM:INTernal:DEPTh<wsp>
<DepthModulation>
Parameter(s)
DepthModulation:
The program data syntax for
<DepthModulation> is defined as a <DECIMAL
NUMERIC PROGRAM DATA> element.
The <DepthModulation> parameter represents
the new dither modulation signal depth in
amperes.
Example(s)
SOUR:PULM:STAT ON
SOUR:WAV:MODE NORMAL
SOUR:PULM:INT:DEPT 0.00255
See Also
SOURce[1..n]:PULM:INTernal:DEPTh?
SOURce[1..n]:PULM:LIMit:DEPTh:HIGH?
SOURce[1..n]:PULM:LIMit:DEPTh:LOW?
SOURce[1..n]:PULM:LIMit:DEPTh:STEP?