Source[1..n]:current:accbase – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 98

SCPI Command Reference
92
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:CURRent:ACCBase
Description
This command is used to set the ACC base
current. The base current is the actual current of
the laser. For this command to function properly,
the instrument must be in High Wavelength
Stability or On/Off modulation mode and the
SOURce[1..n]:STABle? query must return 1
(stable). Normally, this command would rarely
be used.
*RST has no effect on this command.
Syntax
:SOURce[1..n]:CURRent:ACCBase<wsp>
<ACCBase>
Parameter(s)
ACCBase:
The program data syntax for <ACCBase> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <ACCBase> parameter is the desired
current of the instrument in amperes.
Example(s)
SOUR:PULM:STATE OFF
SOUR:WAV:MODE HRES
/* Under FDA jurisdiction, a password must be
entered before turning the laser on. */
SOUR:POW:PROT:SPWD "safekey"
SOUR:POW:STAT ON