Source[1..n]:accoutput – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 92

SCPI Command Reference
86
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:ACCOutput
Description
This command sets the delta temperature and
the delta current of the laser.
At *RST, the delta current and delta temperature
are device-dependent.
Syntax
:SOURce[1..n]:ACCOutput<wsp><DeltaTemp>
,<DeltaCurrent>
Parameter(s)
³
DeltaTemp:
The program data syntax for <DeltaTemp> is
defined as a <DECIMAL NUMERIC PROGRAM
DATA> element.
The <DeltaTemp> parameter represents the
delta temperature of the laser in °C.
The value used must be equal to or between the
values returned by
SOURce[1..n]:TEMPerature:PROTection:LOW
[:LEVel]? and
SOURce[1..n]:TEMPerature:PROTection:HIGH
[:LEVel]?. The query
SOURce[1..n]:TEMPerature:PROTection:STEP?
returns the instrument sensitivity.