Source[1..n]:current:resolution – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual

Page 108

Advertising
background image

SCPI Command Reference

102

IQS-2400

Product-Specific Commands—Description

:SOURce[1..n]:CURRent:RESolution?

Description

This query returns the minimum step available

when changing the laser current with the

SOURce[1..n]:ACCOutput command.

*RST has no effect on this command.

Syntax

:SOURce[1..n]:CURRent:RESolution?

Parameter(s)

None

Response Syntax

<CurrentResolution>

Response(s)

CurrentResolution:
The response data syntax for

<CurrentResolution> is defined as a <NR3

NUMERIC RESPONSE DATA> element.

The <CurrentResolution> response is a value

representing the minimum step available in

amperes.

Example(s)

SOUR:CURR:RES? Returns 1.000000E-004