Calibration[1..n]:value – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 90

SCPI Command Reference
84
IQS-2400
Product-Specific Commands—Description
SOUR:STAB? (repeat until the returned value is
not 1)
SOUR:POW 5.0 DBM
CAL:VAL 5.1DBM
Notes
This command must not be used when the
instrument is in High Wavelength Stability. To
determine the appropriate calibration value,
connect the source output to a power meter. The
power read by the power meter is the parameter
that should be entered with the
CALibration[1..n]:VALue command. To return
the device to its original factory calibration, use
the token DEFault.
The value used must not be more than 10 times
greater or less than 10 times lower than the
factory calibration value. An attempt to exceed
these limits will raise a “Parameter out of range”
error.
See Also
UNIT[1..n]:POWer
SOURce[1..n]:WAVelength:MODE
SOURce[1..n]:PULM:STATe
SOURce[1..n]:POWer:PROTection:SetPassWorD
SOURce[1..n]:POWer:[:LEVel][:IMMediate][:AM
PLitude]
:CALibration[1..n]:VALue