Source[1..n]:pulm:internal: frequency[1|2 – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 132

SCPI Command Reference
126
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:PULM:INTernal:
FREQuency[1|2]?
Description
This query returns the frequency of the internal
dither or On/Off modulation signal. If the
instrument is not in Dither or On/Off modulation
mode, the returned value is undefined.
To get the frequency of the dither mode, the
FREQuency header suffix [1|2] must be 1.
To get the frequency of the On/Off mode, the
FREQuency header suffix [1|2] must be 2.
At *RST, the dither frequency is 1000 Hz and the
On/Off modulation is 10 Hz.
Syntax
:SOURce[1..n]:PULM:INTernal:FREQuency[1|2]?
[<wsp>MAXimum|MINimum|DEFault]
Parameter(s)
Parameter 1:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
MAXimum|MINimum|DEFault.
MINimum is used to retrieve the instrument's
smallest supported value.
MAXimum is used to retrieve the instrument's
greatest supported value.
DEFault is used to retrieve the instrument's
default value.
Response Syntax
<Frequency>