Source[1..n]:pulm:state – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 158

SCPI Command Reference
152
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:PULM:STATe
Description
This command is used to set the modulation to
on or off.
At *RST, modulation is off.
Syntax
:SOURce[1..n]:PULM:STATe<wsp><Modulation
State>
Parameter(s)
ModulationState:
The program data syntax for <ModulationState>
is defined as a <Boolean Program Data>
element. The <ModulationState> special forms
ON and OFF are accepted on input for increased
readability. ON corresponds to 1 and OFF
corresponds to 0.
The <ModulationState> parameter is the
desired modulation state.
0 or OFF, deactivates the modulation.
1 or ON, activates the modulation.
Example(s)
SOUR:PULM:STAT ON
Notes
This command is meant to be used in
conjunction with
SOURce[1..n]:WAVelength:MODE.
When the modulation is on and the resolution
mode is normal, then the dither modulation is
selected. If the resolution mode is high
resolution, then On/Off modulation is selected.
See Also
SOURce[1..n]:PULM:SOURce?
SOURce[1..n]:PULM:SOURce
SOURce[1..n]:PULM:STATe?
SOURce[1..n]:WAVelength:MODE