SCPI Command Reference

WDM Laser Source

83

Product-Specific Commands—Description

Product-Specific Commands—Description

:CALibration[1..n]:VALue

Description

This command is used to set the power

calibration value of the instrument. This

calibration value is used as a correction factor.

At *RST, the calibrated power is set to the factory

value.

Syntax

:CALibration[1..n]:VALue<wsp>

<CalibrationValue[<wsp>DBM|W]>

Parameter(s)

CalibrationValue:
The program data syntax for <CalibrationValue>

is defined as a <DECIMAL NUMERIC PROGRAM

DATA> element followed by an optional

<SUFFIX PROGRAM DATA> element. The

allowed <SUFFIX PROGRAM DATA> elements

are: DBM|W.

The <CalibrationValue> parameter represents

the new calibration value. If you do not specify a

unit, the device will use the current power units

(Watts or dBm). To select the current power units

use the UNIT[1..n]:POWer command.

Example(s)

SOUR:PULM:STAT OFF

SOUR:WAV:MODE NORMAL

/* Under FDA jurisdiction, a password must be

entered before turning the laser on. */

SOUR:POW:PROT:SPWD "safekey"

SOUR:POW:STAT ON