Product-Specific Commands—Description
Product-Specific Commands—Description
:CALibration[1..n]:VALue
Description
This command is used to set the power
calibration value of the instrument. This
calibration value is used as a correction factor.
At *RST, the calibrated power is set to the factory
value.
Syntax
:CALibration[1..n]:VALue<wsp>
<CalibrationValue[<wsp>DBM|W]>
Parameter(s)
CalibrationValue:
The program data syntax for <CalibrationValue>
is defined as a <DECIMAL NUMERIC PROGRAM
DATA> element followed by an optional
<SUFFIX PROGRAM DATA> element. The
allowed <SUFFIX PROGRAM DATA> elements
are: DBM|W.
The <CalibrationValue> parameter represents
the new calibration value. If you do not specify a
unit, the device will use the current power units
(Watts or dBm). To select the current power units
use the UNIT[1..n]:POWer command.
Example(s)
SOUR:PULM:STAT OFF
SOUR:WAV:MODE NORMAL
/* Under FDA jurisdiction, a password must be
entered before turning the laser on. */
SOUR:POW:PROT:SPWD "safekey"
SOUR:POW:STAT ON