Source[1..n]:pulm:internal:shape – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 128

SCPI Command Reference
122
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:PULM:INTernal:SHAPe
Description
This command is used to set the shape of the
internal dither modulation signal. If the
instrument is not in Dither modulation mode, this
command will have no effect.
At *RST, the dither shape is square wave.
Syntax
:SOURce[1..n]:PULM:INTernal:SHAPe<wsp>
SquareWave|TriangleWave
Parameter(s)
DitherModulation:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
SquareWave|TriangleWave.
This parameter is the desired shape of the
internal dither modulation signal.
TriangleWave selects a triangle wave shape.
SquareWave selects a square wave shape.
Example(s)
SOUR:PULM:STAT ON
SOUR:WAV:MODE NORMAL
SOUR:PULM:INT:SHAP SW
Notes
The instrument must be in Dither mode.
See Also
SOURce[1..n]:PULM:INTernal:SHAPe?