Source[1..n]:pulm:limit:depth: delta – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 138

SCPI Command Reference
132
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:PULM:LIMit:DEPTh:
DELTa?
Description
This query returns the minimum step available
when changing the dither modulation signal
depth with the
SOURce[1..n]:PULM:INTernal:DEPTh command.
*RST has no effect on this command.
Syntax
:SOURce[1..n]:PULM:LIMit:DEPTh:DELTa?
Parameter(s)
None
Response Syntax
<DepthPrecision>
Response(s)
DepthPrecision:
The response data syntax for <DepthPrecision>
is defined as a <NR3 NUMERIC RESPONSE
DATA> element.
The <DepthPrecision> response is the
minimum step available for the dither
modulation signal depth in amperes.
Example(s)
SOUR:PULM:LIM:DEPT:DELT? Returns
1.000000E-004
Notes
The step is an indication of the instrument
sensitivity. An attempt to change the dither
modulation signal depth by a quantity less than
the minimum step will be ignored.
See Also
SOURce[1..n]:PULM:INTernal:DEPTh
SOURce[1..n]:PULM:INTernal:DEPTh?
SOURce[1..n]:PULM:LIMit:DEPTh:HIGH?
SOURce[1..n]:PULM:LIMit:DEPTh:LOW?
SOURce[1..n]:PULM:LIMit:DEPTh:STEP?