Source[1..n]:temperature: protection:low[:level – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 163

SCPI Command Reference
WDM Laser Source
157
Product-Specific Commands—Description
:SOURce[1..n]:TEMPerature:
PROTection:LOW[:LEVel]?
Description
This query returns the minimum value at which
the delta temperature can be set with the
SOURce[1..n]:ACCOutput command.
*RST has no effect on this command.
Syntax
:SOURce[1..n]:TEMPerature:PROTection:LOW
[:LEVel]?
Parameter(s)
None
Response Syntax
<LowTemp>
Response(s)
LowTemp:
The response data syntax for <LowTemp> is
defined as a <NR3 NUMERIC RESPONSE DATA>
element.
The <LowTemp> response is the minimum
value at which the delta temperature can be set
in °C. The returned value will be negative.
Example(s)
SOUR:TEMP:PROT:LOW? Returns
-1.000000E+000
See Also
SOURce[1..n]:ACCOutput
SOURce[1..n]:TEMPerature:PROTection:HIGH
[:LEVel]?
SOURce[1..n]:TEMPerature:PROTection:STEP?
SOURce[1..n]:TEMPerature:PROTection:
RESolution?
SOURce[1..n]:TEMPerature:PROTection:DELTa?
SOURce[1..n]:TEMPerature:PROTection:
CHannel?