Source[1..n]:setpoint – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual

Page 160

Advertising
background image

SCPI Command Reference

154

IQS-2400

Product-Specific Commands—Description

:SOURce[1..n]:SETPoint

Description

This command sets the instrument to a

predefined setpoint. No command is provided to

create setpoints. Setpoints have to be created

from within the IQS application.

At *RST, the resolution mode, the wavelength,

the power, the delta temperature and the delta

current are set to the default instrument values.

Syntax

:SOURce[1..n]:SETPoint<wsp>

<SetpointName>,APC|ACC

Parameter(s)

³

SetpointName:

The program data syntax for <SetpointName> is

defined as a <STRING PROGRAM DATA>

element.

The <SetpointName> parameter is the name of

the setpoint. This name is not case-sensitive.

³

LaserMode:

The program data syntax for the second

parameter is defined as a <CHARACTER

PROGRAM DATA> element. The allowed

<CHARACTER PROGRAM DATA> elements for

this parameter are: APC|ACC.

This parameter represents the mode of the

setpoints.

APC sets Normal mode.

ACC sets High Wavelength Stability mode (HWS).

Example(s)

SOUR:SETP "CH38C", APC