Source[1..n]:wavelength:mode – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 179

SCPI Command Reference
WDM Laser Source
173
Product-Specific Commands—Description
:SOURce[1..n]:WAVelength:MODE
Description
This command is used to toggle between the
Normal and High Wavelength Stability (High
Resolution) modes.
At *RST, the resolution mode is normal.
Syntax
:SOURce[1..n]:WAVelength:MODE<wsp>
NORMal|HRESolution
Parameter(s)
SourceMode:
The program data syntax for the first parameter is
defined as a <CHARACTER PROGRAM DATA>
element. The allowed <CHARACTER PROGRAM
DATA> elements for this parameter are:
NORMal|HRESolution.
This parameter represents the desired
wavelength operation mode.
NORMal sets the instrument to Normal mode.
HRESolution sets the instrument to High
Wavelength Stability mode.
Example(s)
SOUR:WAV:MODE HRES
Notes
This command meant to be used in conjunction
with SOURce[1..n]:WAVelength:MODE.
When the modulation is on and the resolution
mode is normal, then the dither modulation is
selected. If the resolution mode is high
resolution, then the On/Off modulation is
selected.
See Also
SOURce[1..n]:WAVelength:MODE?
SOURce[1..n]:PULM:STATe?