Source[1..n]:stable – EXFO IQS-2400 WDM Laser Source for IQS-500/600 User Manual
Page 162

SCPI Command Reference
156
IQS-2400
Product-Specific Commands—Description
:SOURce[1..n]:STABle?
Description
This query indicates whether the source is stable
or stabilizing.
At *RST, the resolution mode, the wavelength,
the power, the delta temperature and the delta
current are set to the default instrument values.
Syntax
:SOURce[1..n]:STABle?
Parameter(s)
None
Response Syntax
<Stable>
Response(s)
Stable:
The response data syntax for <Stable> is
defined as a <NR1 NUMERIC RESPONSE DATA>
element.
The <LimitMax> response is the stabilization
state.
0- The instrument is stabilizing.
1- The instrument has stabilized.
Example(s)
SOUR:STAB?
Notes
After the mode has been changed, the
instrument needs some time to stabilize. It is
very important not to change the mode while the
instrument is stabilizing. This would cause the
instrument to return to its central wavelength in
Normal mode. Before changing the mode, you
should poll SOUR:STAB? until it returns 1,
indicating that the instrument has stabilized.
See Also
SOURce[1..n]:POWer:STATe?