Appendix C-2

Technical Data

Appendix

C

Hydraulics

Hydraulics

Dilution Ratio

When using the dual syringe model equipped with

a 25 mL syringe on the left and a 100 µL syringe on

the right, the dilution ratio is 1/25000, but in order

to stay within the RSD range a maximum ratio of

1/500 is recommended.

Minimum Volume

The minimum programmable transfer volume is 1 µL.

However, precise and accurate transfer of such

volumes depends on the following parameters:
• volumetric capacity of the syringe,

• size of the air gap used,

• geometry of the needle tip,

• flow rate used,

• nature of the liquid to be transferred.

Maximum Volume

The maximum volume that can be aspirated from a

vial is limited to the capacity of the syringe or on the

transfer tubing, whichever is the smaller. The total

aspirated volume (air gaps plus liquid volumes) must

be inferior to this volume.

Flow Rate

Flow rates depend on the nature of the liquid

(viscosity), the capacity of the syringe, and the

internal diameter of tubing. For liquids transferred

via the probe, it is possible to adjust the aspirate and

dispense flow rates according to the nature of the

liquid (e.g. viscous liquids must be aspirated slowly).

The default value for ‘Air flow rate’ is 4 mL/min.

Wathever the capacity of the syringe(s).