5-14

Operation - Advanced Use

5

After completely aspirating the liquid being measured,

you see (for example) the Measurement Screen.

Procedure

1) Aspirate all the liquid contained in the source

vial, plus an air cushion, to ensure that you have

aspirated all the liquid. You should press the

command button to stop aspirating when you

can see the air/liquid interface in the transfer

tubing.

2) Press the Asp key to reverse the direction of

flow; (Asp is replaced by Disp).

3) With the tip of the tubing in the vial, press the

command button to start dispensing the air

cushion.

4) While carefully watching the air/liquid inter-

face, press the command button to stop dispens-

ing at the instant the liquid arrives at the tip of

the tubing

1

.

5) Read the volume shown on the screen (496.7 µL

in the example).

6) With the tip of the tubing in the vial, press the

command button to dispense the liquid back into

the source vial.

7) Before repeating this procedure with other vials,

remember to reverse the direction of flow by

pressing the Disp key again (Disp is replaced by

Asp).

Conclude the current measurement by pressing

STOP or End (see Prime for details).

Titrate

The use of this direct manipulation is the same as

classical titration (using a burette) where the idea is

to run a quantity of reagent into a sample that

includes a pH indicator, up to the point where the

color changes. With 402, you may also use titrate to

fill plates, vials, or bottles.
To titrate, you aspirate a known volume of reagent,

dispense part of the volume required into the sample,

agitate the sample, dispense a small amount,

agitate again, dispense a smaller amount, agitate

again, etc., as you move in finer and finer steps

towards the point where you have added the

precise volume required to change the color.

Software

1

High precision is obtained by

performing this action at a low flow

rate. And, because the piston

accelerates after pressing the

command button, you are advised

to start-and-stop dispensing by

repeatedly pushing the command

button with an increasing frequency

as the liquid nears the tip.