Operation - advanced use, Tasks - detailed descriptions – Gilson 402 User Manual

Page 65

Advertising
background image

5-15

Operation - Advanced Use

5

This is realized using the 402 by pressing the probe’s

command button with increasing rapidity as you

approach the neutralization point. Care must be

taken in selecting the dispense flow rate.
• Press Titrate, Titrate Screen (1) appears. Select

either reservoir or probe as the source using the

Chge soft-key, then press ENTER to confirm your

choice.

• Press the Next soft-key, key in the volume to

aspirate, then press ENTER, Titrate Screen (2)

appears. Normally, you would aspirate sufficient

liquid for several titrations.

• Press the Next soft-key, Titrate Screen (3) appears.
• Follow the instructions given on the screen; you

press the command button on the probe to aspi-

rate and to dispense, when prompted to do so by

the command light on the probe:
- a steady light means you are to aspirate or

dispense liquid,

- a flashing light means you are to aspirate air

via the probe.

Read the volume on the screen, Titrate (4) Screen

(for example), after completing each titration.
Conclude the current titration by pressing STOP or

End (see Prime for details).

Tasks - Detailed Descriptions

Your Method may consist of one or more of the

following tasks. Each task consists of a pre-defined

protocol, from which you select the required mode

of operation, and into which you insert the required

parameter values.

In-run Prompts

For optimization purposes, all volumes and flow

rates, which can be entered as parameters in a task,

can be defined as In-run Prompts. When editing

the task, you key in a value, and then a question

mark (?), the corresponding flashing value has to

be validated or changed before the Method runs. It

is not necessary to re-edit the method to modify the

chosen parameter.

Software

Titrate Screen (1)

Titrate Screen (2)

Titrate Screen (3)

Titrate Screen (4)

Advertising