Operation - advanced use – Gilson 402 User Manual

Page 84

Advertising
background image

5-34

Operation - Advanced Use

5

Gilson

D200 tips are used for volumes from 1 µL to

78 µL (with a 100 µL syringe, minimum recommended

air-push-volume is 20 µL). Gilson

D1000 tips are used

for volumes from 50 µL to 900 µL (minimum

recommended air-push-volume is 100 µL).
The liquid is ejected with an Air push volume of at

least 20 µL when using a Gilson

D200 tip and 100 µL

when using a Gilson

D1000 tip. An automatic

software correction of 2% of the piston stroke

is made to improve accuracy with Gilson

D200 and

D1000 tips. For example, when using a Gilson D200

tip, if an Air push volume of 20 µL is

chosen, the

maximum sample volume will be 78 µL;

this volume

will be handled with the 100 µL syringe.

For the dual

syringe model, the left syringe operates

if the

volume

exceeds 78 µL; for the single syringe model,

an error message is displayed.
A disposable tip should not be fitted, when priming

the 402 in ‘Air Mode’. Droplets of liquid adhering

the tubing after purging must be wiped from the

end, before fitting a tip.
To minimize carry-over, dilution can be performed

with one or more tips. If one tip is used, you use the

Dilute task where the diluent and the sample are

separated by an air gap (volume defined in the

Configuration menu). When several tips are used,

you use multiple runs of the Dispense task to

transfer the liquids, changing the tips where appro-

priate: one to aspirate and dispense the diluent, one

to aspirate and dispense each sample.

Liquid Mode

You can use this mode in the same way as the

standard tip with tapered tubing. However, it is

recommended not to use an air gap, or at least to use

it in a way that keeps the air gap in the disposable-

tip. The use of Gilson

D200 tips does not adversely

affect the performance of the 402.
The Gilson

D200 tip is installed to protect the end

of the probe against corrosive chemicals, and by

changing the tips within a task.
A disposable tip should be fitted, when priming the

402 in ‘Liquid Mode’. The following flow rates are

suggested when priming: small syringes up to 10 mL/

min, larger syringes at least 80 mL/min.

Additional Information