Changing the piston syringes and piston noses, Maintenance – Gilson 402 User Manual

Page 91

Advertising
background image

6-7

Maintenance

6

Changing the Piston Syringes and Piston Noses

Changing the Piston Syringes and
Piston Noses

The piston nose should be replaced after the pump

has performed approximately 30 000 cycles. All

piston noses are replaceable, the process is relatively

easy, but for most pistons (except the 25 mL pistons),

special tools are required to mount the piston nose.

Part Number

Description

100 µL Piston:

F4015062

5 x piston noses, 2 x support collars and mounting tool (for 100 and
250 µL pistons).

F4015072

1 piston assembly kit

250 µL Piston:

F4015063

5 x piston noses and mounting tool (for 100 and 250 µL pistons)

F4015073

1 piston assembly kit

500 µL Piston:

F4015064

5 x piston noses and mounting tool kit

1 mL Piston:

F4015065

5 x 1 mL piston noses and mounting tool

F4015005

1 nose mounting tool

5 mL Piston:

F4015066

1 x piston nose plus O-ring

F4015003

Piston nose mounting kit

10 mL Piston:

F4015067

1 x piston nose plus O-ring

F4015004

Piston nose mounting kit

25 mL Piston:

F4015068

1 x piston nose

Changing the 100 and 250 µL Piston
Noses

Part Number

Description

F4015062

5 x piston noses, 2 x support collars and mounting tool (for 100 and 250 µL pistons)

F4015063

5 x piston noses and mounting tool (for 100 and 250 µL pistons)

To optimize the fitting of the 100 µL and 250 µL

syringes on the syringe pump valve, each syringe is

supplied with a cover seal. The orientation of this

seal must be respected to ensure that the connec-

tion with the syringe is airtight. Figure page 6-10

shows the correct orientation of the cover seal on a

syringe.

Advertising