5-4

Operation - Advanced Use

5

Correct Use of the 402

Aspirating Large Volumes via the
Probe

1. The valve switches to connect the left-

hand syringe to the Tee-junction and

probe.

2. The piston of the right-hand syringe

remains in the top position to prevent

liquid from entering the syringe.

3. The left-hand piston descends to

aspirate the specified volume of sample

liquid.

Aspirating Small Volumes via the
Probe

1. The valve switches to connect the left-

hand syringe to the reservoir, thus

disconnecting the left-hand syringe

from the right syringe.

2. The piston of the left-hand syringe

remains in the top position.

3. The right-hand piston descends to

aspirate the specified volume of sample.

Right-hand Syringe Dispensing to
Probe

1. The valve switches to connect the left-

hand syringe to the reservoir, thus

disconnecting the left-hand syringe

from the right-hand syringe.

2. The piston of the left-hand syringe

remains in the top position.

3. The right-hand piston ascends to

dispense the specified volume of

sample.

(c) Left-hand Syringe
Aspirating from Probe

Left piston
descends
to aspirate

Right piston
remains at
the top

From Tee

From Probe

(d) Right-hand Syringe
Aspirating from Probe

Left piston
remains
inactive

Right piston
descends to
aspirate

From Probe

Valve switched
to Reservoir

(e) Right-hand Syringe
Dispensing to Probe

Valve switched
to Reservoir

To Probe

Left piston
remains
inactive

Right piston
ascends to
aspirate